Comparison of Manual and Fully Automated AIX1000 Rapid Plasma Reagin Assays for Laboratory Diagnosis of Syphilis

手动和全自动AIX1000快速血浆反应素检测法在梅毒实验室诊断中的比较

阅读:1

Abstract

The analytical performance of the AIX1000 system, a fully automated and recently FDA-cleared rapid plasma reagin (RPR) system, was evaluated by comparison to a manual RPR test in a traditional syphilis testing algorithm. A total of 1,028 consecutive serum samples submitted for syphilis testing to the University of North Carolina Hospitals Clinical Immunology Laboratory were tested per each manufacturer's instructions. Among those samples, 996 were nonreactive and 20 were reactive using both the ASI RPR card system and the AIX1000 system. Of the 12 discrepant specimens, 11 were AIX1000 reactive and ASI card nonreactive whereas 1 specimen was ASI card reactive and AIX1000 nonreactive. The sensitivity and specificity of the manual ASI card were 76.0% and 99.8%, respectively, while the sensitivity and specificity of the AIX100 were 100.0% and 99.4%, respectively (sensitivity P = 0.03). Among the 20 concordant reactive specimens, 68.4% of the titers agreed within ±1 dilution between methods. Reproducibility testing of the AIX1000 system demonstrated qualitative and semiquantitative (within ±1 dilution) agreement between specimens tested on different days of 96.0% and 76.0%, respectively, and 100.0% agreement between replicates within the same run. One of 24 samples analyzed under other disease conditions was reactive on both the AIX1000 system and the ASI card. Overall, the fully automated AIX1000 system demonstrated significantly enhanced sensitivity and specificity similar to that of the manual ASI RPR card test, making the AIX1000 system suitable for the laboratory diagnosis of syphilis in a clinical laboratory setting.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。