Conclusions
We show that MSD's R-PLEX NfL assay can reliably quantitate levels of NfL in the CSF and serum from patients with MS and controls, where levels correlate strongly with Simoa.
Methods
NfL was quantitated in samples from 116 individuals from two sites (Ottawa Hospital Research Institute and Mayo Clinic), consisting of patients with MS (n=71) and age- and sex-matched inflammatory neurological controls (n=13) and non-inflammatory controls (n=32). Correlation of NfL levels between CSF and serum was assessed in paired samples in a subset of MS patients and controls (n=61). Additionally, we assessed the correlation between NfL levels obtained with MSD's R-PLEX® and Quanterix's single molecule array (Simoa®) assays in CSF and serum (n=32).
Results
Using the R-PLEX, NfL was quantitated in 99% of the samples tested, and showed a broad range in the CSF (82-500,000 ng/L) and serum (8.84-2,014 ng/L). Nf-L levels in both biofluids correlated strongly (r=0.81, p<0.0001). Lastly, Nf-L measured by MSD's R-PLEX and Quanterix's Simoa assays were highly correlated for both biofluids (CSF: r=0.94, p<0.0001; serum: r=0.95, p<0.0001). Conclusions: We show that MSD's R-PLEX NfL assay can reliably quantitate levels of NfL in the CSF and serum from patients with MS and controls, where levels correlate strongly with Simoa.