Fast reversed-phase liquid chromatography to reduce back exchange and increase throughput in H/D exchange monitored by FT-ICR mass spectrometry

快速反相液相色谱法可减少反向交换并提高氢/氘交换的通量,并通过傅里叶变换离子回旋共振质谱法进行监测。

阅读:1

Abstract

In solution-phase hydrogen/deuterium exchange (HDX), it is essential to minimize the back-exchange level of H for D after the exchange has been quenched, to accurately assign protein conformation and protein-protein or protein-ligand interactions. Reversed-phase HPLC is conducted at low pH and low temperature to desalt and separate proteolytic fragments. However, back exchange averages roughly 30% because of the long exposure to H(2)O in the mobile phase. In this report, we first show that there is no significant backbone amide hydrogen back exchange during quench and digestion; backbone exchange occurs primarily during subsequent liquid chromatography separation. We then show that a rapid reversed-phase separation reduces back exchange for HDX by at least 25%, resulting from the dramatically reduced retention time of the peptide fragments on the column. The influence of retention time on back exchange was also evaluated. The rapid separation coupled with high-resolution FT-ICR MS at 14.5 T provides high amino acid sequence coverage, high sample throughput, and high reproducibility and reliability.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。