Metal wear-induced pseudotumour following an endoprosthetic knee replacement for Ewing sarcoma

尤文氏肉瘤行膝关节置换术后金属磨损引起的假瘤

阅读:1

Abstract

Pseudotumours are well recognised as a complication of metal-on-metal hip arthroplasties and are thought to develop on the basis of an innate and adaptive immune response to cobalt-chrome (Co-Cr) wear particles. We report a case of a large pseudotumour that developed following a knee endoprosthetic replacement (EPR) undertaken for Ewing sarcoma. The lesion contained necrotic and degenerate connective tissue in which there were numerous scattered metal wear-containing macrophages, eosinophil polymorphs, lymphocytes, plasma cells and aseptic lymphocyte-dominated vascular-associated lesion-like lymphoid aggregates. Metal ion levels were elevated. No evidence of infection or tumour was noted and it was concluded that the lesion was most likely an inflammatory pseudotumour developing on the basis of an innate and adaptive immune response to components of Co-Cr metal wear derived from the knee EPR.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。