Functional Evaluation of the Reusable JuniorSTAR® Half-Unit Insulin Pen

可重复使用的 JuniorSTAR® 半单位胰岛素笔的功能评估

阅读:1

Abstract

BACKGROUND: The functional performance of the JuniorSTAR(®) (Sanofi, Paris, France) half-unit insulin pen was evaluated through a series of specific objective tests to assess the dose accuracy, pen weight, injection force, and dialing torque. METHOD: Pens (n = 60) were tested under standard atmospheric conditions with 3 different types of insulins manufactured by Sanofi (insulin glargine, insulin glulisine, and biphasic insulin isophane). The dose accuracy was tested according to the ISO 11608-1:2012 standards. Injection doses of 0.010, 0.155, and 0.300 ml were evaluated. For mean weight evaluation, the pens without the cartridge were weighed on precision balances. The injection force was measured using a texture analyzer and the dialing torque was measured using a torque meter. RESULTS: JuniorSTAR met the ISO 11608-1:2012 criteria for dose accuracy as all the delivered doses were within the predefined limits for all types of insulin tested. The mean weight of the JuniorSTAR pen was 33.4 g (SD = 0.075). The mean injection force was 6.0 N (SD = 0.8), 4.3 N (SD = 0.4), and 5.1 N (SD = 0.6) for insulin glargine, insulin glulisine, and biphasic insulin isophane, respectively. The mean dialing torque was 5.09 Ncm (SD = 0.29) and 5.88 Ncm (SD = 0.53) for setting and correcting a dose, respectively. CONCLUSIONS: Together with results from a previously reported usability survey, these results show that the JuniorSTAR reusable, half-unit pen is a lightweight and accurate device for insulin delivery with a dialing torque and injection force suitable for young people with type 1 diabetes.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。