An optimized system for interventional magnetic resonance imaging-guided stereotactic surgery: preliminary evaluation of targeting accuracy

一种用于介入磁共振成像引导立体定向手术的优化系统:靶向精度的初步评估

阅读:1

Abstract

BACKGROUND: Deep brain stimulation electrode placement with interventional magnetic resonance imaging (MRI) has previously been reported using a commercially available skull-mounted aiming device (Medtronic Nexframe MR) and native MRI scanner software. This first-generation method has technical limitations that are inherent to the hardware and software used. A novel system (SurgiVision ClearPoint) consisting of an aiming device (SMARTFrame) and software has been developed specifically for interventional MRI, including deep brain stimulation. OBJECTIVE: To report a series of phantom and cadaver tests performed to determine the capability, preliminary accuracy, and workflow of the system. METHODS: Eighteen experiments using a water phantom were used to determine the predictive accuracy of the software. Sixteen experiments using a gelatin-filled skull phantom were used to determine targeting accuracy of the aiming device. Six procedures in 3 cadaver heads were performed to compare the workflow and accuracy of ClearPoint with Nexframe MR. RESULTS: Software prediction experiments showed an average error of 0.9 ± 0.5 mm in magnitude in pitch and roll (mean pitch error, -0.2 ± 0.7 mm; mean roll error, 0.2 ± 0.7 mm) and an average error of 0.7 ± 0.3 mm in X-Y translation with a slight anterior (0.5 ± 0.3 mm) and lateral (0.4 ± 0.3 mm) bias. Targeting accuracy experiments showed an average radial error of 0.5 ± 0.3 mm. Cadaver experiments showed a radial error of 0.2 ± 0.1 mm with the ClearPoint system (average procedure time, 88 ± 14 minutes) vs 0.6 ± 0.2 mm with the Nexframe MR (average procedure time, 92 ± 12 minutes). CONCLUSION: This novel system provides the submillimetric accuracy required for stereotactic interventions, including deep brain stimulation placement. It also overcomes technical limitations inherent in the first-generation interventional MRI system.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。