Catheter-based photoacoustic endoscope

导管式光声内窥镜

阅读:8
作者:Joon-Mo Yang, Chiye Li, Ruimin Chen, Qifa Zhou, K Kirk Shung, Lihong V Wang

Abstract

We report a flexible shaft-based mechanical scanning photoacoustic endoscopy (PAE) system that can be potentially used for imaging the human gastrointestinal tract via the instrument channel of a clinical video endoscope. The development of such a catheter endoscope has been an important challenge to realize the technique's benefits in clinical settings. We successfully implemented a prototype PAE system that has a 3.2-mm diameter and 2.5-m long catheter section. As the instrument's flexible shaft and scanning tip are fully encapsulated in a plastic catheter, it easily fits within the 3.7-mm diameter instrument channel of a clinical video endoscope. Here, we demonstrate the intra-instrument channel workability and in vivo animal imaging capability of the PAE system.