Feasibility Study of Using XRV-124 Scintillation Detector for Collinearity Measurement in Uniform Scanning Proton Therapy

利用XRV-124闪烁探测器进行均匀扫描质子治疗共线测量的可行性研究

阅读:1

Abstract

PURPOSE: The purpose of this work is to study the feasibility of using an XRV-124 scintillation detector in measuring the collinearity of the x-ray system and uniform scanning proton beam. METHODS: A brass aperture for Snout 10 was manufactured. The center of the aperture had an opening of 1 cm in diameter (4 cm for the film measurements). The 2D kV x-ray images of the XRV-124 were acquired such that the marker inside the detector is aligned to the imaging isocenter. After obtaining the optimal camera settings, a uniform scanning proton beam was delivered for various ranges (12 g/cm(2) to 28 g/cm(2) in step size of 2 g/cm(2)). For each range, 10 monitor units (MU) of the first layer were delivered to the XRV-124 detector. Collinearity tests were repeated by using EDR2 and EBT3 films following our current quality assurance protocol in practice. The results from the XRV-124 measurements were compared against the collinearity results from EDR2 and EBT3 films. RESULTS AND DISCUSSION: The collinearity results were evaluated in the horizontal (x) and vertical (y) directions. The average deviation in collinearity in the x-direction was -0.24 ± 0.30 mm, 0.57 ± 0.39 mm, and -0.27 ± 0.14 mm for EDR2, EBT3, and XRV-124, respectively. In the y-direction, the average deviation was 0.39 ± 0.07 mm, 0.29 ± 0.14 mm, and 0.39 ± 0.03 mm for EDR2, EBT3, and XRV-124, respectively. CONCLUSION: The measurement results from the XRV-124 and films are in good agreement. Compared to film, the use of the XRV-124 detector for collinearity measurements in uniform scanning protons is more efficient and provides results in real time.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。