A High Aspect Ratio Bifurcated 128-Microchannel Microfluidic Device for Environmental Monitoring of Explosives

一种用于爆炸物环境监测的高纵横比分叉128微通道微流控装置

阅读:1

Abstract

Design and evolution of explosives monitoring and detection platforms to address the challenges of trace level chemical identification have led investigations into the use of intricately designed microfluidic devices. Microfluidic devices are unique tools that possess distinct characteristics that, when designed properly and configured with optical and fluidic components, can produce detection platforms with unmatched performance levels. Herein, we report the design, fabrication and integration of a bifurcated high aspect ratio microfluidic device containing 128 microchannels (40 mm × 40 μm × 250 μm; L × W × H) for explosives detection at trace levels. Aspect ratios measuring >6:1 support improved receptor-target molecule interactions, higher throughput and extremely low limits of detection (LOD). In addition to superior assay sensitivity, the bifurcated microfluidic device provides greater durability and versatility for substrate modification. Using the explosive 2,4,6-trinitrotoluene (TNT) as the model compound in a fluorescence-based displacement immunoassay, we report LODs for TNT at 10 parts-per-trillion (pptr) using a neutravidin-coated biotinylated anti-TNT microfluidic device. Solution to wall interactions were also simulated in COMSOL Multiphysics to understand fluid flow characteristics. Reynolds numbers were calculated to be 0.27⁻2.45 with a maximum pressure of 1.2 × 10(-2) psi.