Currents recorded through small areas of squid axon membrane with an internal virtual ground voltage clamp

利用内部虚拟地电压钳记录鱿鱼轴突膜小区域的电流

阅读:1

Abstract

A new voltage-clamp apparatus for the squid axon has been implemented to enable recording of currents through small areas of axon membrane. The performance of this clamp was tested by recording total sodium currents from perfused axons (I total) and sodium currents from small membrane patches (I patch), which were recorded from inside the axon with an L-shaped pipette. The I patch records, although four orders of magnitude smaller than I total, were stable and showed normal kinetics and voltage dependence, and appeared to reflect the activation of a small population of normal sodium channels. The size of the current recorded from the patch was mainly a function of the tip diameter of the L-shaped pipette and of the shunt resistance between inside the pipette and the axoplasm.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。