Robotic inhibition assay for determination of HMG-CoA reductase inhibitors in human plasma

利用机器人抑制试验测定人血浆中HMG-CoA还原酶抑制剂

阅读:1

Abstract

The cholesterol-lowering drug simvastatin (SIMV, Zocor reduced heart attacks by 42% in patients who had high cholesterol levels and suffered from heart disease. Upon oral administration, SIMV is quickly hydrolyzed to its beta-hydroxyacid and other acid metabolites, which are potent inhibitors of 3-hydroxy-3-methylglutaryl coenzyme A reductase. A Tecan-based enzyme inhibition assay has been developed to improve the existing Zymark-based assay for the determination of both active and total concentrations of HMG-CoA reductase inhibitors in human plasma. A Tecan Genesis 200 robotic workstation equipped with eight probes and customized hardware was utilized to achieve higher sample throughput and improve assay reproducibility and mechanical stability. The developed enzyme inhibition assay was validated over two concentration ranges of 0.4-20 ng equivalent/mL, and 2-50 ng equivalent/mL. Intra- and interday precision data (coefficient of variation (CV)) for both concentration ranges were less than 9%, with an accuracy of 93-107%. The interday precision for the determination of quality control (QC) samples was less than 2% and 8%, respectively. The respective interday QC accuracy values were 93-103% and 97-104%. Good linearity across the two concentration ranges was observed, with acceptable reproducibility. This improved enzyme inhibition assay has been utilized to analyze human plasma samples from several clinical studies.