Pre-clinical dosimetry of a new six-channel applicator for high-dose-rate treatment of esophageal cancer

用于食管癌高剂量率治疗的新型六通道施药器的临床前剂量测定

阅读:1

Abstract

PURPOSE: To evaluate the dosimetry of a six-channel high-dose-rate (HDR) applicator for treatment of esophageal cancer with respect to lateral directionality and heterogeneous media. MATERIAL AND METHODS: A computed tomography (CT)- and magnetic resonance imaging (MRI)-compatible esophageal applicator consisting of 2 inflatable portions (anchor and therapeutic balloons) with 6 longitudinal treatment catheters equally spaced circumferentially was constructed. Treatment plans were prepared using Oncentra 4.5 for various catheter loadings and target locations and sizes. Calculated dose distributions were compared to measured distributions obtained using film and a water phantom. Balloon inflations with water and with air were tested. RESULTS: TG-43 dose calculations matched measurements well when inflation balloons were filled with water. When air was used to inflate, model-based dose calculations (TG-186) improved the comparison with measurement. Several cases with simulated ring targets demonstrated better dose conformity to non-uniform targets compared to a single central catheter. Additionally, the use of this applicator compared to a single catheter, gave rise to considerable improvement in sparing non-target tissue. CONCLUSIONS: Lateral dose modulation is achievable with the applicator described in this work. The use of TG-186 dose calculation made a small improvement in heterogeneous media.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。