Real-time intraoperative 4D full-range FD-OCT based on the dual graphics processing units architecture for microsurgery guidance

基于双图形处理单元架构的实时术中4D全范围频域光学相干断层扫描(FD-OCT)技术,用于显微外科手术导航

阅读:1

Abstract

Real-time 4D full-range complex-conjugate-free Fourier-domain optical coherence tomography (FD-OCT) is implemented using a dual graphics processing units (dual-GPUs) architecture. One GPU is dedicated to the FD-OCT data processing while the second one is used for the volume rendering and display. GPU accelerated non-uniform fast Fourier transform (NUFFT) is also implemented to suppress the side lobes of the point spread function to improve the image quality. Using a 128,000 A-scan/second OCT spectrometer, we obtained 5 volumes/second real-time full-range 3D OCT imaging. A complete micro-manipulation of a phantom using a microsurgical tool is monitored by multiple volume renderings of the same 3D date set with different view angles. Compared to the conventional surgical microscope, this technology would provide the surgeons a more comprehensive spatial view of the microsurgical site and could serve as an effective intraoperative guidance tool.