Contactless continuous heart rate monitoring system using ballistocardiography

采用弹道心动描记法的非接触式连续心率监测系统

阅读:1

Abstract

Cardiovascular disease is the number one cause of death in the world and is a serious problem. In the case of cardiopulmonary arrest due to myocardial infarction, the survival rate is as low as 13.3% one month after resuscitation, which birthed the need for continuous heart monitoring. In this study, we develop a Ballistocardiogram (BCG) measurement system using a load cell installed on a chair and a heart rate estimation algorithm that is robust to waveform changes, with the aim of constructing a non-contact heart rate acquisition system. The proposed system was evaluated by utilizing data obtained from 13 healthy subjects and 1 subject with abnormal ECG who were simultaneously measured with ECG. The output of the BCG system was confirmed to change with the same period as the ECG data obtained as the correct answer, and the synchronization of the R-peak positions was confirmed for all cases. As a result of comparing the heart rate intervals estimated from BCG and those obtained from ECG, it was confirmed that the same heart rate variability (HRV) features could be obtained even for abnormal ECG subject.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。