"Carryover phenomenon" in subcutaneous implantable cardioverter-defibrillator: Residual energy caused early shock to subsequent ventricular tachycardia

皮下植入式心脏复律除颤器中的“残留现象”:早期电击导致的残余能量引发后续室性心动过速

阅读:1

Abstract

We reported a case of early S-ICD shock due to the carryover of previously charged energy. Depending on the interval with the preceding event, an early shock may occur for the subsequent event. Especially, in cases where non-sustained VTs occur frequently, the indication for S-ICD surgery should be considered carefully.