Comparing hepatic 2D and 3D magnetic resonance elastography methods in a clinical setting - Initial experiences

临床环境下肝脏二维和三维磁共振弹性成像方法的比较——初步经验

阅读:1

Abstract

PURPOSE: Continuous monitoring of liver fibrosis progression in patients is not feasible with the current diagnostic golden standard (needle biopsy). Recently, magnetic resonance elastography (MRE) has emerged as a promising method for such continuous monitoring. Since there are different MRE methods that could be used in a clinical setting there is a need to investigate whether measurements produced by these MRE methods are comparable. Hence, the purpose of this pilot study was to evaluate whether the measurements of the viscoelastic properties produced by 2D (stiffness) and 3D (elasticity and 'G abs,Elastic') MRE are comparable. MATERIALS AND METHODS: Seven patients with diffuse or suspect diffuse liver disease were examined in the same day with the two MRE methods. 2D MRE was performed using an acoustic passive transducer, with a 1.5 T GE 450 W MR system. 3D MRE was performed using an electromagnetic active transducer, with a 1.5 T Philips Achieva MR system. Finally, mean viscoelastic values were extracted from the same anatomical region for both methods by an experienced radiologist. RESULTS: Stiffness correlated well with the elasticity, R (2) = 0.96 (P < 0.001; slope = 1.08, intercept = 0.61 kPa), as well as with 'G abs,Elastic' R (2) = 0.96 (P < 0.001; slope = 0.95, intercept = 0.28 kPa). CONCLUSION: This pilot study shows that different MRE methods can produce comparable measurements of the viscoelastic properties of the liver. The existence of such comparable measurements is important, both from a clinical as well as a research perspective, since it allows for equipment-independent monitoring of disease progression.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。