Patella-Femoral Joint Revision Surgery Following a Rare and Interesting Case of Implant Failure: A Case Report and Retrospective Analysis of Our Hospitals Implant Data

髌股关节翻修手术:一例罕见且有趣的植入物失效病例报告及我院植入物数据的回顾性分析

阅读:3

Abstract

According to the national joint registry, patello-femoral joint (PFJ) replacement accounts for 1% of all knee arthroplasty procedures in the UK, 1014 of which were performed in 2018. The femoro-patella vialla (FPV) implant by MicroPort orthopaedics has a high reported rate of revision, more than four times that of other knee replacements. The mechanisms of failure are usually loss of fixation at the bone-implant interface. We observe a rare and unusual case form of patella component failure whereby the facets of the implant remain imbedded within the patella as a result of shear force, leaving a loose patella button within the knee. A literature review was conducted of reported modes of failure of the FPV implant and retrospective analysis of our units' results looking into rate of revision, modes of failure, and assessment of a single surgeon's patient outcome measure (Knee Society Scores). Retrospective analysis of our hospital's data revealed that 11% of FPV implants were revised in our unit, a large percentage of which were due to pain and progressive osteoarthritis. We observed revision rates of 5.84% at 3 years, 10.21% at 5 years, and 13.13% at 10 years, which is favourable to national joint registry revision estimates of 6.99%, 10.13%, and 19.10%, respectively. Knee Society Scores reveal a mean improvement of 39.65 (-38 to 100) at follow-up following FPV unicompartmental knee arthroplasty surgery. Our unit demonstrated better results from patient outcome measures when compared to literature and a lower rate of revision when compared to national figures.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。