The Difference in the Assessment of Knee Extension/Flexion Angles during Gait between Two Calibration Methods for Wearable Goniometer Sensors

两种可穿戴测角仪传感器校准方法在步态过程中膝关节伸展/屈曲角度评估方面的差异

阅读:1

Abstract

Frontal and axial knee motion can affect the accuracy of the knee extension/flexion motion measurement using a wearable goniometer. The purpose of this study was to test the hypothesis that calibrating the goniometer on an individual's body would reduce errors in knee flexion angle during gait, compared to bench calibration. Ten young adults (23.2 ± 1.3 years) were enrolled. Knee flexion angles during gait were simultaneously assessed using a wearable goniometer sensor and an optical three-dimensional motion analysis system, and the absolute error (AE) between the two methods was calculated. The mean AE across a gait cycle was 2.4° (0.5°) for the on-body calibration, and the AE was acceptable (<5°) throughout a gait cycle (range: 1.5-3.8°). The mean AE for the on-bench calibration was 4.9° (3.4°) (range: 1.9-13.6°). Statistical parametric mapping (SPM) analysis revealed that the AE of the on-body calibration was significantly smaller than that of the on-bench calibration during 67-82% of the gait cycle. The results indicated that the on-body calibration of a goniometer sensor had acceptable and better validity compared to the on-bench calibration, especially for the swing phase of gait.