Preclinical In-Vivo Evaluation and Screening of Zinc Based Degradable Metals for Endovascular Stents

血管内支架用锌基可降解金属的临床前体内评估与筛选

阅读:10
作者:Roger J Guillory 2nd, Alexander A Oliver, Emma Davis, Elisha J Earley, Jaroslaw W Drelich, Jeremy Goldman

Abstract

Zinc alloy development and characterization for vascular stent application has been facilitated by many standardized and inexpensive methods. In contrast, overly simplistic in vitro approaches dominate the preliminary biological testing of materials. In 2012, our group introduced a metal wire implantation model in rats as a cost effective and realistic approach for the biocompatibility evaluation of degradable materials in the vascular environment. Here, we have adapted metrics routinely used for evaluating stents to quantitatively characterize the long-term progression of the neointima that forms around zinc based wire implants. Histological cross-sections were used to measure the length of neointimal protrusion from the wire into the lumen (denoted wire to lumen thickness), the base neointimal length (describing the breadth of neointimal activation), and the neointimal area. These metrics were used to provide in depth characterization details for neointimal responses to Zn-Mg and Zn-Li alloys and may be used to compare different materials.