Electrical methods are among the primarily studied non-invasive glucose measurement techniques; however, various factors affect the accuracy of the sensors used. Of these, the temperature is a critical factor; hence, the effects of temperature on the electrical properties of blood components are investigated in this study. Furthermore, the changes in the electrical properties of blood according to the glucose level are corrected by considering the effects of temperature on the electrical properties. An impedance sensor is developed and used to measure whole blood impedance in 10 healthy participants at various temperatures and glucose levels. Subsequently, the conductivities of the plasma and cytoplasm were extracted. Changes in the electrical properties of the blood components are then analyzed using linear regression and repeated measures ANOVA. The electrical conductivities of plasma and cytoplasm increased with increasing temperatures (plasma: 0.0397 (slope), 0.7814 (R(2)), cytoplasm: 0.014 (slope), 0.694 (R(2))). At three values of increasing glucose levels (85.4, 158.1, and 271.8 mg/dL), the electrical conductivities of the plasma and cytoplasm decreased. These tendencies are more significant upon temperature corrections (p-values; plasma: 0.001, 0.001, cytoplasm: 0.003, 0.002). The relationships between temperature and electrical conductivity changes can thus be used for temperature corrections in blood glucose measurement.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。