Evaluation of micro Electroretinograms Recorded with Multiple Electrode Array to Assess Focal Retinal Function

利用多电极阵列记录微视网膜电图来评估局部视网膜功能

阅读:16
作者:Momo Fujii, Genshiro A Sunagawa, Mineo Kondo, Masayo Takahashi, Michiko Mandai

Abstract

Full-field electroretinograms (ERGs) are used to objectively assess the mass function of the retina, whereas focal ERGs are used to evaluate the focal retinal function. The purpose of this study was to determine the usefulness of a multiple electrode array (MEA) system for recording ex vivo micro ERGs (mERGs) together with multiunit spike responses of the retinal ganglion cells (RGCs) to assess focal retinal function in isolated mouse retinas. The a- and b-waves of the full-field ERGs were present in the mERG. The b-wave was blocked by L-AP4, an inhibitor of the mGluR6 receptor, and the OFF-component was blocked by exposure to PDA, an antagonist of ionotropic glutamate receptors, with a corresponding RGC responses. mERGs were also recorded from mice with progressive retinal degeneration, the C57BL/6J-Pde6b(rd1-2J)/J (rd1) mice, from which conventional full-field ERGs are non-recordable. A blockade of the glutamate receptors indicated that the negative wave of rd1 mice do not originate from the photoreceptors but from the second or third order neurons. This technique of recording mERGs will be useful in assessing the focal properties of the retinas obtained from eyes with pathology and also to follow the recovery of the physiology of the retina in regenerative studies.