Correction technique for cascade gammas in I-124 imaging on a fully-3D, Time-of-Flight PET Scanner

全三维飞行时间PET扫描仪上I-124成像中级联γ射线的校正技术

阅读:1

Abstract

It has been shown that I-124 PET imaging can be used for accurate dose estimation in radio-immunotherapy techniques. However, I-124 is not a pure positron emitter, leading to two types of coincidence events not typically encountered: increased random coincidences due to non-annihilation cascade photons, and true coincidences between an annihilation photon and primarily a coincident 602 keV cascade gamma (true coincidence gamma-ray background). The increased random coincidences are accurately estimated by the delayed window technique. Here we evaluate the radial and time distributions of the true coincidence gamma-ray background in order to correct and accurately estimate lesion uptake for I-124 imaging in a time-of-flight (TOF) PET scanner. We performed measurements using a line source of activity placed in air and a water-filled cylinder, using F-18 and I-124 radio-isotopes. Our results show that the true coincidence gamma-ray backgrounds in I-124 have a uniform radial distribution, while the time distribution is similar to the scattered annihilation coincidences. As a result, we implemented a TOF-extended single scatter simulation algorithm with a uniform radial offset in the tail-fitting procedure for accurate correction of TOF data in I-124 imaging. Imaging results show that the contrast recovery for large spheres in a uniform activity background is similar in F-18 and I-124 imaging. There is some degradation in contrast recovery for small spheres in I-124, which is explained by the increased positron range, and reduced spatial resolution, of I-124 compared to F-18. Our results show that it is possible to perform accurate TOF based corrections for I-124 imaging.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。