Validation of Treatment Planning Systems based on Dose Calculation Accuracy for Extended SSD Total Body Irradiation with and without Beam Spoiler

基于剂量计算精度的扩展源皮距全身照射治疗计划系统验证(有/无束流扰流器)

阅读:1

Abstract

AIMS: This study aimed to confirm the accuracy of the dose calculation by commercially available treatment planning system (TPS) at the extended SSD = 350 cm with and without beam spoiler configurations. MATERIALS AND METHODS: Two extended SSD plans, one without a beam spoiler and another with a beam spoiler, were created on a virtual phantom in Eclipse TPSs for 6 MV photon beam. Using dosimetry tools, percentage depth doses (PDDs) and in-line profiles of both plans were generated. Beam properties of the TPS calculated PDDs and profiles were compared with manually calculated and measured PDDs and profiles. Reference depth dose was calculated in two different phantoms and measured using an ion chamber and TLDs. RESULTS: At the extended SSD setup, the TPS calculated and measured PDD with and without beam spoiler were in agreement within 2%. This agreement was observed at the surface and in the buildup region as well. Flatness and symmetry of the profiles with and without beam spoiler were within ± 3%. The TPS calculated dose at a reference depth in a homogeneous and a heterogeneous phantom was agreed with measured dose using an ion-chamber and TLDs. CONCLUSION: This study validated the TPS for extended SSD total body irradiation (TBI) dose calculation with and without beam spoiler. Developing computed tomography based field in field plan for homogeneous dose distribution may be feasible. This technique may allow us to treat extended SSD TBI in less time while maintaining dose calculation accuracy.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。