Chest-MRI under pulsatile flow ventilation: A new promising technique

脉流通气下胸部磁共振成像:一项前景广阔的新技术

阅读:1

Abstract

OBJECTIVES: Magnetic resonance imaging (MRI) of the chest has long suffered from its sensitivity to respiratory and cardiac motion with an intrinsically low signal to noise ratio and a limited spatial resolution. The purpose of this study was to perform chest MRI under an adapted non invasive pulsatile flow ventilation system (high frequency percussive ventilation, HFPV®) allowing breath hold durations 10 to 15 times longer than other existing systems. METHODS: One volunteer and one patient known for a thymic lesion underwent a chest MRI under ventilation percussion technique (VP-MR). Routinely used sequences were performed with and without the device during three sets of apnoea on inspiration. RESULTS: VP-MR was well tolerated in both cases. The mean duration of the thoracic stabilization was 10.5 min (range 8.5-12) and 5.8 min (range 5-6.2) for Volunteer 1 and Patient 1, respectively. An overall increased image quality was seen under VP-MR with a better delineation of the mediastinal lesion for Patient 1. Nodules discovered in Volunteer 1 were confirmed with low dose CT. CONCLUSION: VP-MR was feasible and increased spatial resolution of chest MRI by allowing acquisition at full inspiration during thoracic stabilization approaching prolonged apnoea. This new technique could be of benefit to numerous thoracic disorders.