High spatial resolution identification of hematoma in inhomogeneous head phantom using broadband fNIR system

利用宽带近红外光谱系统对非均匀头部模型中的血肿进行高空间分辨率识别

阅读:1

Abstract

This paper presents a novel method for early detection of hematomas using highly sensitive optical fNIR imaging methods based on broadband photon migration. The NIR experimental measurements of inhomogeneous multi-layer phantoms representing human head are compared to 3D numerical modeling over broadband frequencies of 30-1000 MHz. A finite element method (FEM) simulation of the head phantom are compared to measurements of insertion loss and phase using custom-designed broadband free space optical transmitter (Tx) and receiver (Rx) modules that are developed for photon migration at wavelengths of 670 nm, 795 nm, 850 nm, though results of 670 nm are discussed here. Standard error is used to compute error between 3D FEM modeling and experimental measurements by fitting experimental data to the [Formula: see text]. Error results are shown at narrowband and broadband frequency modulation in order to have confidence in 3D numerical modeling. A novel method is established here to identify presence of hematoma based on first and second derivatives of changes in insertion loss and phase (∆IL and ∆IP), where frequency modulated photons sensitive to different sizes of hematoma is identified for wavelength of 670 nm. The high accuracy of this comparison provides confidence in optical bio-imaging and its eventual application to TBI detection.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。