System Accuracy Assessment of a Combined Invasive and Noninvasive Glucometer

侵入式和非侵入式联合血糖仪的系统准确性评估

阅读:1

Abstract

BACKGROUND: The pain associated with pricking the fingertip for blood glucose self-testing is considered to be a major burden in diabetes treatment. This study was performed to evaluate the system accuracy of the invasive TensorTip Combo Glucometer (CoG) device component in accordance with ISO15197:2015 requirements and to explore the accuracy of the noninvasive tissue glucose prediction component. METHODS: One hundred samples were obtained from people with type 1 and type 2 diabetes and healthy volunteers (43 females, 57 males; age: 53 ± 16 years), with glucose distribution as requested by the ISO standard. Three strip lots were tested twice by healthcare professionals in comparison to YSI 2300 Stat Plus reference method followed by a noninvasive tissue glucose reading (NI-CoG). Mean Absolute (Relative) Difference (MARD) was calculated and a consensus error grid (CEG) analysis was performed. RESULTS: The ISO system accuracy criteria were met with the invasive strip technology by 586/600 of the data points (97.1%) and for each strip lot separately. All invasive results (100%) were within CEG-zone A and total MARD was calculated to be 7.1%. With the noninvasive reading, 99% of raw data points were in A + B (91.1% and 7.8%), and the total MARD was calculated to be 18.1%. DISCUSSION: The invasive component of the CoG device was shown to be in full compliance with the current ISO15197 criteria. Good results were also obtained with the NI-CoG tissue glucose prediction. This noninvasive technology would potentially be suitable for frequent pain-free glucose monitoring in many people with diabetes.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。