Meeting the New FDA Standard for Accuracy of Self-Monitoring Blood Glucose Test Systems Intended for Home Use by Lay Users

符合美国食品药品监督管理局 (FDA) 关于家用血糖自我监测测试系统准确性的新标准

阅读:1

Abstract

The OneTouch Verio Reflect blood glucose monitor (BGM) has market clearance in several countries based in part on fulfilling the lay user and system accuracy criteria described in ISO15197:2015. However, the Food and Drug Administration (FDA) does not recognize the accuracy criteria in ISO15197 as a basis for gaining regulatory clearance for these devices. The current study evaluates the BGM using the accuracy guidelines issued by the agency for self-monitoring blood glucose test systems for over-the-counter use. Glucose results were accurate vs comparator over a wide glucose range and met lay user and glucose accuracy criteria at extreme glucose values as described in the FDA guidance.Clinicaltrials.gov NCT03851549.