Analysis of "Performance of a Factory-Calibrated, Real-Time Continuous Glucose Monitoring System in Pediatric Participants With Type 1 Diabetes"

对“工厂校准的实时连续血糖监测系统在1型糖尿病儿科患者中的性能分析”

阅读:1

Abstract

Accuracy of sensors play an important role in the acceptance and long-term use of CGM which is related to improved glycemic outcome. This has been lower in children and adolescents in the past for reasons such as size of the device, pain, and necessity to calibrate sensors but also inaccuracy and perceived nonreliability on alarms among others. In the study of Welsh et al, performance of a new, factory-calibrated sensor was assessed. The accuracy of the sensor, as measured in MARD, paired values within 20/20% or 15/15% and surveillance error grid analysis showed very good results, although less so in low value range <70 mg/dL, falling rate of change, and on first day of use. Accurate glucose measurements especially in these incidences are of utmost importance to people with diabetes as their treatment decisions are based on these, also in the beginning of sensor use.