Validation of a laser-assisted wound measurement device for measuring wound volume

验证一种用于测量伤口体积的激光辅助伤口测量装置

阅读:1

Abstract

BACKGROUND: Accurate and precise wound measurement is an essential part of the medical record when monitoring a patient with a chronic wound. This study was designed to determine if a new device, a laser-assisted wound measurement (LAWM) device, provides valid measurements for wound area, depth, and volume. METHODS: We compared four methods to evaluate the area and volume of 12 wounds of differing size and depth that were created on the dorsum of a sacrificed pig. We evaluated the LAWM device, digital photograph assessment with National Institutes of Health ImageJ software, measurements of depth with a ruler, and weight-to-volume assessment with dental paste. We then sought to cross validate this data with further analyses obtained from these measurements using a Play-Doh®-based wound as a model for constant area with different depths. RESULTS: We demonstrate that the LAWM device measures wound area accurately. Depth (and therefore volume) measurements, however, are artificially low. This inaccuracy is the same for shallow and deep wounds. CONCLUSIONS: The inaccuracy in the depth and volume measurements with the LAWM device results in an artificially low measurement. However, this may not affect percentage difference measurements. Further studies will need to be performed to determine if this device can accurately determine wound changes in the clinical setting.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。