Validity and Reliability of the Newly Developed Surface Electromyography Device for Measuring Muscle Activity during Voluntary Isometric Contraction

新型表面肌电图仪测量自主等长收缩期间肌肉活动的有效性和可靠性研究

阅读:1

Abstract

OBJECTIVE: The purpose of this study was to establish the validity and reliability of the newly developed surface electromyography (sEMG) device (PSL-EMG-Tr1) compared with a conventional sEMG device (BTS-FREEEMG1000). METHODS: In total, 20 healthy participants (10 males, age 30.3 ± 2.9 years; 10 females, age 22.3 ± 2.7 years) were recruited. EMG signals were recorded simultaneously on two devices during three different isometric contractions (maximal voluntary isometric contraction (MVIC, 40% MVIC, 80% MVIC)). Two trials were performed, and the same session was repeated after 1 week. EMG amplitude recorded from the dominant biceps brachii (BB) and rectus femoris (RF) muscles was analyzed for reliability using intrasession intraclass correlation coefficient (ICC). Concurrent validity of the two devices was determined using Pearson's correlation coefficient. RESULTS: Nonnormalized sEMG data showed moderate to very high reliability for all three contraction levels (ICC = 0.832-0.937 (BB); ICC = 0.814-0.957 (RF)). Normalized sEMG values showed no to high reliability (ICC = 0.030-0.831 (BB); ICC = 0.547-0.828 (RF)). sEMG signals recorded by the PSL-EMG-Tr1 showed good to excellent validity compared with the BTS-FREEEMG1000, at 40% MVIC (r = 0.943 (BB), r = 0.940 (RF)) and 80% MVIC (r = 0.983 (BB); r = 0.763 (RF)). CONCLUSIONS: The PSL-EMG-Tr1 was performed with acceptable validity. Furthermore, the high accessibility and portability of the device are useful in adjusting the type and intensity of exercise.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。