Comparison of automated and manual protocols for magnetic resonance imaging assessment of liver iron concentration

比较自动和手动磁共振成像评估肝脏铁浓度的方法

阅读:1

Abstract

OBJECTIVE: To compare automated and manual magnetic resonance imaging protocols for estimating liver iron concentrations at 1.5 T. MATERIALS AND METHODS: Magnetic resonance imaging examination of the liver was performed in 53 patients with clinically suspected hepatic iron overload and in 21 control subjects. Liver iron concentrations were then estimated by two examiners who were blinded to the groups. The examiners employed automated T2* and T1 mapping, as well as manual T2* and signal-intensity-ratio method. We analyzed accuracy by using ROC curves. Interobserver and intraobserver agreement were analyzed by calculating two-way intraclass correlation coefficients. RESULTS: The area under the ROC curve (to discriminate between patients and controls) was 0.912 for automated T2* mapping, 0.934 for the signal-intensity-ratio method, 0.908 for manual T2*, and 0.80 for T1 mapping, the last method differing significantly from the other three. The level of interobserver and intraobserver agreement was good (intraclass correlation coefficient, 0.938-0.998; p < 0.05). Correlations involving T1 mapping, although still significant, were lower. CONCLUSION: At 1.5 T, T2* mapping is a rapid tool that shows promise for the diagnosis of liver iron overload, whereas T1 mapping shows less accuracy. The performance of T1 mapping is poorer than is that of T2* methods.