Locking plate osteosynthesis for a femoral fracture and subsequent nonunion in a patient with osteopetrosis

股骨骨折伴骨硬化症患者发生锁定钢板内固定术后出现骨不连

阅读:1

Abstract

INTRODUCTION: Osteopetrosis is a skeletal disorder characterized by increased osteodensity and a remodeling defect. The fragility of dense sclerotic bones may lead to an increased incidence of fractures. Although internal fixation can be performed, technical challenges may be experienced because of the increased bone density. Complications such as delayed union, nonunion, or implant failure may occur postoperatively. PRESENTATION OF CASE: We describe a patient with autosomal-dominant osteopetrosis type 2 who suffered a shaft fracture below a plate of his right femur. We performed osteosynthesis with a single locking plate. Union was delayed, and plate breakage occurred along with nonunion of the fracture. The nonunion was addressed using double locking plates, which secured fixation and allowed complete fracture healing. DISCUSSION: There were three reasons of nonunion in our case. First, we left gaps between the fragments. Second, we used mainly cerclage wires, rather than screws, for plate fixation, which led to inadequate stability. Third, the patient was large (height 167 cm, weight 93.1 kg), so the single plate provided insufficient fixing force. We then used double locking plates and attained stronger internal fixation with complete fracture healing. CONCLUSION: Double plating with locking plates may be an effective treatment option for femoral fractures in patients with osteopetrosis.