Lack of sensitivity of an IVD/CE-labelled kit targeting the S gene for detection of SARS-CoV-2

针对SARS-CoV-2 S基因的体外诊断/CE认证试剂盒灵敏度不足

阅读:1

Abstract

OBJECTIVES: New molecular tests for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) are being rapidly launched in response to the coronavirus disease 2019 (COVID-19) pandemic. The aim of this study was to evaluate the analytical and clinical performance of the VIASURE SARS-CoV-2 S gene RT-PCR Kit on the BD Max™ system and to compare results with those obtained with the cobas® SARS-CoV-2 test on the cobas® 6800 system. METHODS: For testing the analytical performance, reference material was used. Clinical samples (n = 101) obtained from individuals with symptoms compatible with COVID-19 were studied. Oropharyngeal and nasopharyngeal swabs were collected by using either ESwab™ or UTM™ collection systems. RESULTS: When the analytical performance was evaluated, the sample containing the lowest SARS-CoV-2 concentration tested negative with the VIASURE test whereas results obtained with the cobas® test were found to be concordant with the results expected. Six out of the 101 clinical samples (5.9%) showed an inhibition with the VIASURE test. When analysing the remaining 95 clinical samples, 27 were found to be negative with both assays. Of 68 samples that were positive with the cobas® test, the VIASURE test missed 21 (30.9 %) samples. All of those 21 samples had shown Ct values ≥ 31 with the cobas® 6800 system. None of the samples tested positive with the VIASURE test and negative with the cobas® test. CONCLUSIONS: The VIASURE test was impaired by a lack of sensitivity and a relatively high number of invalid results. When using the VIASURE test for routine testing, a significant number of COVID-19-positive samples would have been missed.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。