Correlation of objective image quality and working length measurements in different CBCT machines: An ex vivo study

不同锥形束CT机客观图像质量与工作长度测量值的相关性:一项离体研究

阅读:1

Abstract

To investigate potential correlations between objective CBCT image parameters and accuracy in endodontic working length determination ex vivo. Contrast-to-noise ratio (CNR) and spatial resolution (SR) as fundamental objective image parameters were examined using specific phantoms in seven different CBCT machines. Seven experienced observers were instructed and calibrated. The order of the CBCTs was randomized for each observer and observation. To assess intra-operator reproducibility, the procedure was repeated within six weeks with a randomized order of CBCT images. Multivariate analysis (MANOVA) did not reveal any influence of the combined image quality factors CNR and SR on measurement accuracy. Inter-operator reproducibility as assessed between the two observations was poor, with a mean intra-class correlation (ICC) of 0.48 (95%-CI  0.38, 0.59) for observation No. 1. and 0.40 (95%-CI 0.30, 0.51) for observation No. 2. Intra-operator reproducibility pooled over all observers between both observations was only moderate, with a mean ICC of 0.58 (95%-CI 0.52 to 0.64). Within the limitations of the study, objective image quality measures and exposure parameters seem not to have a significant influence on accuracy in determining endodontic root canal lengths in CBCT scans. The main factor of variance is the observer.