Comparison of HIV-1 RNA measurements obtained by using plasma and dried blood spots in the automated abbott real-time viral load assay

比较使用血浆和干血斑样本在雅培全自动实时病毒载量检测中测得的HIV-1 RNA值。

阅读:1

Abstract

Dried blood spots (DBS) may be a promising alternative specimen type to plasma for measuring the viral load (VL) in HIV-infected individuals in resource-limited settings. However, characterization of assay performance using DBS is incomplete. In this prospective study, the VL was measured in parallel using plasma and DBS specimens collected at the same time from 157 HIV-1-infected individuals. DBS were prepared by dispensing 50 μl of blood onto filter paper cards and were stored desiccated at -20°C. Nucleic acid extraction from plasma and DBS was performed automatically using the Abbott m2000sp instrument, and the VL was measured using the RealTime HIV-1 VL assay, which has a lower limit of detection of 40 HIV RNA copies/ml. The correlation between plasma and DBS results was good (R = 0.91; P < 0.001). The mean difference in the VL (DBS minus plasma) was 0.35 log copies (standard deviation [SD], 0.47 log copies). A total of 40 (26%) paired specimens had a difference of >0.5 log copy, and in 12 (7.8%) it was >1 log copy. the VL from DBS was measurable in 95.7% of specimens with a plasma VL of >2.74 log copies (550 HIV RNA copies/ml). In summary, the VL can reliably be measured using DBS with the Abbott RealTime HIV-1 assay. The estimated lower limit of detection of this automated methodology on DBS is 550 copies/ml, a threshold that may be acceptable for periodic VL monitoring in patients on antiretroviral therapy in resource-limited settings, where early detection of virologic treatment failure is often problematic.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。