Test-retest reliability of a functional electromechanical dynamometer on swing eccentric hamstring exercise measures in soccer players

功能性机电测力计在足球运动员摆动离心腘绳肌运动测量中的重测信度

阅读:1

Abstract

BACKGROUND: The use of a functional electromechanical dynamometer (FEMD) has been proposed as a valid and effective tool to evaluate specific movement patterns. The aim of this study was to determine the reliability of FEMD on swing eccentric hamstring exercise (SEHE) measures in soccer players. METHODS: Nineteen federated male soccer players (20.74 ± 4.04 years) performed the SEHE at three different isokinetic velocities (20-40-60 cm/s). These evaluations were conducted in four sessions, two for familiarization and two for registration. The average and maximum load (N) of the three isokinetic velocities was calculated from the values obtained from the FEMD (Dynasystem(®), Bangalore). RESULTS: The main results of this research showed that the reliability was high for the average load in the condition of 40 cm/s, presenting the highest ICC value (0.94). For maximum load, reliability was high in the condition of 20 cm/s. The manifestation of the most reliable load was the maximum load (ICC = 0.91-0.87). CONCLUSIONS: FEMD (Dynasystem(®), Bangalore) is a reliable device to evaluate the eccentric strength of the hamstring muscles in soccer players.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。