Abstract

REVIVE SE (REVIVE) is a closed-ended, self-expanding stent retriever used in the RIVER JAPAN study. We present our early experience with REVIVE for revascularization of acute ischemic stroke (AIS) in patients who have failed or are ineligible for intravenous recombinant tissue plasminogen activator treatment. This prospective, single-arm, non-randomized, multicenter registry study followed up patients undergoing mechanical thrombectomy with REVIVE for 90 days. The primary endpoint was a post-procedure Thrombolysis in Cerebral Infarction (TICI) score ≥2a. Secondary endpoints were clot migration/embolization; recanalization without symptomatic intracranial hemorrhage (ICH) at 24 h; symptomatic ICH; good neurological outcome (modified Rankin Scale score ≤2 National Institute of Health Stroke Scale (NIHSS) score decrease ≥10) at day 90; device- or procedure-related serious adverse events (SAEs) and mortality at day 90. To confirm non-inferiority of REVIVE, results were compared with historical data of the Merci Retriever. About 49 patients were enrolled (median age 73 years; males 46.9%; middle cerebral artery (MCA) occlusion 83.7%; median NIHSS score 17). A post-procedure TICI score ≥2a was observed in 73.5% (36/49, 95% confidence interval [CI] 58.9-85.1) of patients. No post-procedural clot migration/embolization events occurred. Successful recanalization without symptomatic ICH was observed in 62.5% (30/48, 95% CI 47.4-76.0). The good neurological outcome was achieved in 66.7% (32/48) patients. Symptomatic ICH and device- or procedure-related SAEs were reported in 6.3% and 12.2% of patients, respectively. Two deaths were reported. REVIVE demonstrated equivalent efficacy and safety as the Merci Retriever. Results suggest that REVIVE is effective and safe in recanalizing occluded intracranial arteries in AIS.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。