B-mode ultrasound assessment of pupillary function: Feasibility, reliability and normal values

B型超声评估瞳孔功能:可行性、可靠性和正常值

阅读:1

Abstract

PURPOSE: To evaluate B-mode ultrasound as a novel method for the examination of pupillary function and to provide normal values for the pupillary reflex as assessed by B-mode ultrasound. METHODS: 100 subjects (49 female, 51 male, mean [range] age 51 [18-80 years]) with no history of ophthalmologic disease, no clinically detectable pupillary defects, and corrected visual acuity ≥ 0.8 were included in this prospective observational study. B-mode ultrasound was performed with the subjects eyes closed using an Esaote-Mylab25 system according to current guidelines for orbital insonation. A standardized light stimulus was applied. RESULTS: The mean ± standard deviation left and right pupillary diameters (PD) at rest were 4.7 ± 0.8 and 4.5 ± 0.8 mm. Following an ipsilateral light stimulus (Lstim), left and right constricted PD were 2.8 ± 0.6 and 2.7 ± 0.6 mm. Following a contralateral Lstim, left and right constricted PD were 2.7 ± 0.6 and 2.6 ± 0.5 mm. Left and right pupillary constriction time (PCT) following ipsilateral Lstim were 970 ± 261.6 and 967 ± 220 ms. Left and right PCT following a contralateral Lstim were 993.8 ± 192.6 and 963 ± 189.4 ms. Patient age was inversely correlated with PD at rest and with PD after ipsilateral and contralateral Lstim (all p<0.001), but not with PCT. CONCLUSIONS: B-mode ultrasound is a simple, rapid and objective method for the quantitative assessment of pupillary function, which may prove useful in a variety of settings where eyelid retraction is impeded or an infrared pupillometry device is unavailable.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。