Dose indices in dental cone beam CT and correlation with dose-area product

牙科锥形束CT的剂量指数及其与剂量面积乘积的相关性

阅读:1

Abstract

OBJECTIVES: In the 2011 project "Safety and efficacy of a new and emerging dental X-ray modality (SEDENTEXCT)", it was suggested that dose index (DI) and dose-area product (DAP) could be easily measured and used as diagnostic reference levels (DRLs), which would help in the management of radiation doses to patients in optimum exposure settings. Such indices could be directly related to effective dose. The purposes of this study, therefore, were to measure and calculate the DI and DAP in cone beam CT (CBCT) machines and to evaluate the correlation between the two. METHODS: Dose measurements were performed on three-dimensional cone beam CT (3D-CBCT) machines [3D Accuitomo (J. Morita Mfg. Corp., Kyoto, Japan), Veraviewepocs (J. Morita Mfg. Corp.) and CS9300 (Carestream, New York, NY)] by exposing a cylindrical poly-methyl methacrylate (PMMA) phantom using a CT ionization chamber. These dose measurements were used for the calculation of Dose Indices 1 and 2, according to the methodology suggested by SEDENTEXCT. The DAP was measured using a DAP meter that was attached to the detector to cover the entire irradiated area. RESULTS: The DI1 ranged from 53.6 mR to 216.6 mR, the DI2 ranged from 77.1 mR to 325.0 mR and the DAP ranged from 101.1 mGy cm(2) to 457.9 mGy cm(2), depending on the machines and exposure settings. Index 2 had a better correlation with the DAP than Index 1. CONCLUSIONS: The DIs and DAP proposed by SEDENTEXCT may be useful for establishing DRLs for dental CBCT machines; however, further studies are necessary to determine which of these indices provide accurate dose estimates proportionally relating to the effective dose.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。