Internal lead shielding for clinical electron treatments

用于临床电子治疗的内部铅屏蔽

阅读:1

Abstract

Electron beams are often used to treat superficial lesions of the lip, cheek, nose, and ear. Lead is frequently used to block distal structures. It is customary to place an internal bolus of low atomic number in between the tissue and the lead to reduce electron backscatter from the lead. Space for the lead and the internal bolus is quite limited. A previous method for estimating the thickness of the lead plus internal bolus is not self-consistent and leads to a larger than necessary thickness. A new method is described here to provide a quick, accurate, and self-consistent estimate of the minimum necessary thickness of the internal bolus and the lead for incident electron beam energies of 4, 6, 8, 9, and 10 MeV as a function of the thickness of the overlying tissue. This method limits the dose enhancement at the tissue/bolus interface due to the underlying lead to 10%. Measurements made with gafchromic film validate this methodology.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。