Cardiac motion and its dosimetric impact during radioablation for refractory ventricular tachycardia

心脏运动及其在难治性室性心动过速放射消融术中的剂量学影响

阅读:1

Abstract

INTRODUCTION: Cardiac radioablation (CR) is a noninvasive treatment option for patients with refractory ventricular tachycardia (VT) during which high doses of radiation, typically 25 Gy, are delivered to myocardial scar. In this study, we investigate motion from cardiac cycle and evaluate the dosimetric impact in a cohort of patients treated with CR. METHODS: This retrospective study included eight patients treated at our institution who had respiratory-correlated and ECG-gated 4DCT scans acquired within 2 weeks of CR. Deformable image registration was applied between maximum systole (SYS) and diastole (DIAS) CTs to assess cardiac motion. The average respiratory-correlated CT (AVG(resp) ) was deformably registered to the average cardiac (AVG(cardiac) ), SYS, and DIAS CTs, and contours were propagated using the deformation vector fields (DVFs). Finally, the original treatment plan was recalculated on the deformed AVG(resp) CT for dosimetric assessment. RESULTS: Motion magnitudes were measured as the mean (SD) value over the DVFs within each structure. Displacement during the cardiac cycle for all chambers was 1.4 (0.9) mm medially/laterally (ML), 1.6 (1.0) mm anteriorly/posteriorly (AP), and 3.0 (2.8) mm superiorly/inferiorly (SI). Displacement for the 12 distinct clinical target volumes (CTVs) was 1.7 (1.5) mm ML, 2.4 (1.1) mm AP, and 2.1 (1.5) SI. Displacements between the AVG(resp) and AVG(cardiac) scans were 4.2 (2.0) mm SI and 5.8 (1.4) mm total. Dose recalculations showed that cardiac motion may impact dosimetry, with dose to 95% of the CTV dropping from 27.0 (1.3) Gy on the AVG(resp) to 20.5 (7.1) Gy as estimated on the AVG(cardiac) . CONCLUSIONS: Cardiac CTV motion in this patient cohort is on average below 3 mm, location-dependent, and when not accounted for in treatment planning may impact target coverage. Further study is needed to assess the impact of cardiac motion on clinical outcomes.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。