Error on the stopping power ratio of ERKODENT's mouthpiece for head and neck carbon ion radiotherapy treatment

ERKODENT公司用于头颈部碳离子放射治疗的口含器的阻止本领比误差

阅读:2

Abstract

The errors on the stopping power ratio (SPR) of mouthpiece samples from ERKODENT were evaluated. Erkoflex and Erkoloc-pro from ERKODENT and samples that combined Erkoflex and Erkoloc-pro were computed tomography (CT)-scanned using head and neck (HN) protocol at the East Japan Heavy Ion Center (EJHIC), and the values were averaged to obtain the CT number. The integral depth dose of the Bragg curve with and without these samples was measured for 292.1, 180.9, and 118.8 MeV/u of the carbon-ion pencil beam using an ionization chamber with concentric electrodes at the horizontal port of the EJHIC. The average value of the water equivalent length (WEL) of each sample was obtained from the difference between the range of the Bragg curve and the thickness of the sample. To calculate the difference between the theoretical and measured values, the theoretical CT number and SPR value of the sample were calculated using the stoichiometric calibration method. Compared with the Hounsfield unit (HU)-SPR calibration curve used at the EJHIC, the SPR error on each measured and theoretical value was calculated. The WEL value of the mouthpiece sample had an error of approximately 3.5% in the HU-SPR calibration curve. From this error, it was evaluated that for a mouthpiece with a thickness of 10 mm, a beam range error of approximately 0.4 mm can occur, and for a mouthpiece with a thickness of 30 mm, a beam range error of approximately 1 mm can occur. For a beam passing through the mouthpiece in HN treatment, it would be practical to consider a mouthpiece margin of 1 mm to avoid beam range errors if ions pass through the mouthpiece.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。