Intrafractional accuracy and efficiency of a surface imaging system for deep inspiration breath hold during ablative gastrointestinal cancer treatment

消融性胃肠道癌症治疗中深吸气屏气时表面成像系统的分次内准确性和效率

阅读:1

Abstract

PURPOSE: Beam gating with deep inspiration breath hold (DIBH) usually depends on some external surrogate to infer internal target movement, and the exact internal movement is unknown. In this study, we tracked internal targets and characterized residual motion during DIBH treatment, guided by a surface imaging system, for gastrointestinal cancer. We also report statistics on treatment time. METHODS AND MATERIALS: We included 14 gastrointestinal cancer patients treated with surface imaging-guided DIBH volumetrically modulated arc therapy, each with at least one radiopaque marker implanted near or within the target. They were treated in 25, 15, or 10 fractions. Thirteen patients received treatment for pancreatic cancer, and one underwent separate treatments for two liver metastases. The surface imaging system monitored a three-dimensional surface with ± 3 mm translation and ± 3° rotation threshold. During delivery, a kilovolt image was automatically taken every 20° or 40° gantry rotation, and the internal marker was identified from the image. The displacement and residual motion of the markers were calculated. To analyze the treatment efficiency, the treatment time of each fraction was obtained from the imaging and treatment timestamps in the record and verify system. RESULTS: Although the external surface was monitored and limited to ± 3 mm and ± 3°, significant residual internal target movement was observed in some patients. The range of residual motion was 3-21 mm. The average displacement for this cohort was 0-3 mm. In 19% of the analyzed images, the magnitude of the instantaneous displacement was > 5 mm. The mean treatment time was 17 min with a standard deviation of 4 min. CONCLUSIONS: Precaution is needed when applying surface image guidance for gastrointestinal cancer treatment. Using it as a solo DIBH technique is discouraged when the correlation between internal anatomy and patient surface is limited. Real-time radiographic verification is critical for safe treatments.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。