The influence of beam optics asymmetric distribution on dose in scanning carbon-ion radiotherapy

扫描碳离子放射治疗中光束光学不对称分布对剂量的影响

阅读:1

Abstract

PURPOSE: To quantify the influence of beam optics asymmetric distribution on dose. METHODS: Nine reference cubic targets and corresponding plans with modulation widths (M) of 3, 6, and 9 cm and with center depths (CDs) of 6, 12, and 24 cm were generated by the treatment planning system (TPS). The Monte Carlo code FLUKA was used for simulating the dose distribution from the aforementioned original plans and the dose perturbation by varying ±5%, ±15%, ±20%, ±25%, and ±40% in spot full width half maximum to the X-direction while keeping consistent in the Y-direction. The dosimetric comparisons in dose deviation, γ-index analysis, lateral penumbra, and flatness were evaluated. RESULTS: The largest 3D absolute mean deviation was 15.0% ± 20.9% (mean ± standard deviation) in M3CD6, whereas with the variation from -15% to +20%, the values were below 5% for all cube plans. The lowest 2D γ-index passing rate was 80.6% with criteria of 2%-2 mm by a +40% variation in M3CD6. For the M9CD24 with a -40% variation, the maximum 1D dose deviations were 5.6% and 15.7% in the high-dose region and the edge of the radiation field, respectively. The maximum deviations of penumbra and flatness were 3.4 mm and 11.4%, respectively. CONCLUSIONS: The scenario of beam optics asymmetric showed relatively slight influence on the global dose distribution but severely affected dose on the edge of the radiation field. For scanning carbon-ion therapy facilities, beam spot lateral profile settings in TPS base data should be properly handled when beam optics asymmetry variation is over 15%.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。