Determining the longitudinal accuracy and reproducibility of T(1) and T(2) in a 3T MRI scanner

确定3T磁共振成像扫描仪中T(1)和T(2)的纵向准确性和可重复性

阅读:1

Abstract

PURPOSE: To determine baseline accuracy and reproducibility of T(1) and T(2) relaxation times over 12 months on a dedicated radiotherapy MRI scanner. METHODS: An International Society of Magnetic Resonance in Medicine/National Institute of Standards and Technology (ISMRM/NIST) System Phantom was scanned monthly on a 3T MRI scanner for 1 year. T(1) was measured using inversion recovery (T(1) -IR) and variable flip angle (T(1) -VFA) sequences and T(2) was measured using a multi-echo spin echo (T(2) -SE) sequence. For each vial in the phantom, accuracy errors (%bias) were determined by the relative differences in measured T(1) and T(2) times compared to reference values. Reproducibility was measured by the coefficient of variation (CV) of T(1) and T(2) measurements across monthly scans. Accuracy and reproducibility were mainly assessed on vials with relaxation times expected to be in physiological ranges at 3T. RESULTS: A strong linear correlation between measured and reference relaxation times was found for all sequences tested (R(2 ) > 0.997). Baseline bias (and CV[%]) for T(1) -IR, T(1) -VFA and T(2) -SE sequences were +2.0% (2.1), +6.5% (4.2), and +8.5% (1.9), respectively. CONCLUSIONS: The accuracy and reproducibility of T(1) and T(2) on the scanner were considered sufficient for the sequences tested. No longitudinal trends of variation were deduced, suggesting less frequent measurements are required following the establishment of baselines.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。