Dosimetry with a clinical linac adapted to FLASH electron beams

利用适用于 FLASH 电子束的临床直线加速器进行剂量测定

阅读:1

Abstract

PURPOSE: To assess dosimetric properties and identify required updates to commonly used protocols (including use of film and ionization chamber) pertaining to a clinical linac configured into FLASH (ultra-high dose rate) electron mode. METHODS: An 18MV photon beam of a Varian iX linac was converted to FLASH electron beam by replacing the target and the flattening filter with an electron scattering foil. The dose was prescribed by entering the MUs through the console. Fundamental beam properties, including energy, dose rate, dose reproducibility, field size, and dose rate dependence on the SAD, were examined in preparation for radiobiological experiments. Gafchromic EBT-XD film was evaluated for usability in measurements at ultra-high dose rates by comparing the measured dose to the inverse square model. Selected previously reported models of chamber efficiencies were fitted to measurements in a broad range of dose rates. RESULTS: The performance of the modified linac was found adequate for FLASH radiobiological experiments. With exception of the increase in the dose per MU on increase in the repetition rate, all fundamental beam properties proved to be in line with expectations developed with conventional linacs. The field size followed the theorem of similar triangles. The highest average dose rate (2 × 10(4)  Gy/s) was found next to the internal monitor chamber, with the field size of FWHM = 1.5 cm. Independence of the dose readings on the dose rate (up to 2 × 10(4)  Gy/s) was demonstrated for the EBT-XD film. A model of recombination in an ionization chamber was identified that provided good agreement with the measured chamber efficiencies for the average dose rates up to at least 2 × 10(3)  Gy/s. CONCLUSION: Dosimetric measurements were performed to characterize a linac converted to FLASH dose rates. Gafchromic EBT-XD film and dose rate-corrected cc13 ionization chamber were demonstrated usable at FLASH dose rates.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。