Impact of setup errors on multi-isocenter volumetric modulated arc therapy for craniospinal irradiation

设置误差对多等中心容积调强弧形治疗颅脊髓照射的影响

阅读:1

Abstract

Multi-isocenter volumetric modulated arc therapy (VMAT) is recommended for craniospinal irradiation (CSI) to smooth the dose distribution in the junction regions relying solely on inverse optimization. However, few studies have measured the dosimetric impact of setup errors on this multi-isocenter VMAT in the junction areas. The purpose of this study is to evaluate the impact of positional errors during VMAT CSI with two-dimension (2D) and three-dimension (3D) dosimetric measurements. A total of 20 patients treated by three-isocenter VMAT CSI were retrospectively reviewed and analyzed. A 3D diode array ArcCHECK and radiochromic film EBT3 were applied to measure the percentage gamma passing rates (%GPs) and dose distributions in the junction areas between the cranial/upper-spinal and the upper/lower-spinal fields with intentionally introduced setup errors of ± 1 mm, ±2 mm, ±3 mm, ±5 mm, and ± 8 mm, respectively. The length and volume of planning target volume (PTV) for these CSI patients ranged from 50.14 to 80.8 cm, and 1572.3 to 2114.5 cm(3) , respectively. The %GPs for ±3 mm, ±5 mm, and ±8 mm positional errors were around 95%, 90%, and 85%, respectively, in the junction areas. The dosimetric verification results with EBT3 films indicated that cold and hot areas were observed with the increase of introduced setup errors. In conclusion, the dosimetric verification with intentionally introduced setup errors demonstrated that positional errors within 3 mm have a little impact for VMAT CSI, although setup errors should be minimized. Relying on the inverse optimization of VMAT to smooth the dose distribution in the junction areas is feasible for CSI.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。