Development and validation of a comprehensive patient-specific quality assurance program for a novel stereotactic radiation delivery system for breast lesions

针对新型乳腺病灶立体定向放射治疗系统,开发并验证一套全面的患者特异性质量保证方案

阅读:1

Abstract

PURPOSE: The GammaPod is a dedicated prone breast stereotactic radiosurgery (SRS) machine composed of 25 cobalt-60 sources which rotate around the breast to create highly conformal dose distributions for boosts, partial-breast irradiation, or neo-adjuvant SRS. We describe the development and validation of a patient-specific quality assurance (PSQA) system for the GammaPod. METHODS: We present two PSQA methods: measurement based and calculation based PSQA. The measurements are performed with a combination of absolute and relative dose measurements. Absolute dosimetry is performed in a single point using a 0.053-cc pinpoint ionization chamber in the center of a polymethylmethacrylate (PMMA) breast phantom and a water-filled breast cup. Relative dose distributions are verified with EBT3 film in the PMMA phantom. The calculation-based method verifies point doses with a novel semi-empirical independent-calculation software. RESULTS: The average (± standard deviation) breast and target sizes were 1263 ± 335.3 cc and 66.9 ± 29.9 cc, respectively. All ion chamber measurements performed in water and the PMMA phantom agreed with the treatment planning system (TPS) within 2.7%, with average (max) difference of -1.3% (-1.9%) and -1.3% (-2.7%), respectively. Relative dose distributions measured by film showed an average gamma pass rate of 97.0 ± 3.2 when using a 3%/1 mm criteria. The lowest gamma analysis pass rate was 90.0%. The independent calculation software had average agreements (max) with the patient and QA plan calculation of 0.2% (2.2%) and -0.1% (2.0%), respectively. CONCLUSION: We successfully implemented the first GammaPod PSQA program. These results show that the GammaPod can be used to calculate and deliver the predicted dose precisely and accurately. For routine PSQA performed prior to treatments, the independent calculation is recommended as it verifies the accuracy of the planned dose without increasing the risk of losing vacuum due to prolonged waiting times.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。