Retrospective quality metrics review of stereotactic radiosurgery plans treating multiple targets using single-isocenter volumetric modulated arc therapy

回顾性质量指标评估采用单中心容积调强弧形治疗治疗多靶点立体定向放射外科计划

阅读:1

Abstract

PURPOSE: To characterize key plan quality metrics in multi-target stereotactic radiosurgery (SRS) plans treated using single-isocenter volumetric modulated arc therapy (VMAT) in comparison to dynamic conformal arc (DCA) plans treating single target. To investigate the feasibility of quality improvement in VMAT planning based on previous planning knowledge. MATERIALS AND METHODS: 97 VMAT plans of multi-target and 156 DCA plans of single-target treated in 2017 at a single institution were reviewed. A total of 605 targets were treated with these SRS plans. The prescription dose was normalized to 20 Gy in all plans for this analysis. Two plan quality metrics, target conformity index (CI) and normal tissue volume receiving more than 12 Gy (V12Gy), were calculated for each target. The distribution of V12Gy per target was plotted as a function of the target volume. For multi-target VMAT plans, the number of targets being treated in the same plan and the distance between targets were calculated to evaluate their impact on V12Gy. VMAT plans that had a large deviation of V12Gy from the average level were re-optimized to determine the possibility of reducing the variation of V12Gy in VMAT planning. RESULTS: Conformity index of multi-target VMAT plans were lower than that of DCA plans while the mean values of 12 Gy were comparable. The V12Gy for a target in VMAT plan did not show apparent dependence on the total number of targets or the distance between targets. The distribution of V12Gy exhibited a larger variation in VMAT plans compared to DCA plans. Re-optimization of outlier plans reduced V12 Gy by 33.9% and resulted in the V12Gy distribution in VMAT plans more closely resembling that of DCA plans. CONCLUSION: The benchmark data on key plan quality metrics were established for single-isocenter multi-target SRS planning. It is feasible to use this knowledge to guide VMAT planning and reduce high V12Gy outliers.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。