Couch modeling optimization for tomotherapy planning and delivery

用于断层放射治疗计划和实施的床位模型优化

阅读:2

Abstract

We sought to validate new couch modeling optimization for tomotherapy planning and delivery. We constructed simplified virtual structures just above a default setting couch through a planning support system (MIM Maestro, version 8.2, MIM Software Inc, Cleveland, OH, USA). Based on ionization chamber measurements, we performed interactive optimization and determined the most appropriate physical density of these virtual structures in a treatment planning system (TPS). To validate this couch optimization, Gamma analysis and these statistical analyses between a three-dimensional diode array QA system (ArcCHECK, Sun Nuclear, Melbourne, FL, USA) results and calculations from ionization chamber measurements were performed at 3%/2 mm criteria with a threshold of 10% in clinical QA plans. Using a virtual model consisting of a center slab density of 4.2 g/cm(3) and both side slabs density of 1.9 g/cm(3) , we demonstrated close agreement between measured dose and the TPS calculated dose. Agreement was within 1% for all gantry angles at the isocenter and within 2% in off-axis plans. In validation of the couch modeling in a clinical QA plan, the average gamma passing rate improved approximately 0.6%-5.1%. It was statistically significant (P < 0.05) for all treatment sites. We successfully generated an accurate couch model for a TomoTherapy TPS by interactively optimizing the physical density of the couch using a planning support system. This modeling proved to be an efficient way of correcting the dosimetric effects of the treatment couch in tomotherapy planning and delivery.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。